Skip to content

Certifying a medical wearable: where to start

Guide, medical wearable

A wearable with a medical purpose, a watch that detects an arrhythmia, a glucose-monitoring patch, a connected blood-pressure cuff, stacks an unusual number of certification regimes. It is at once a medical device, medical electrical equipment, a software product, radio equipment and a battery product, each with its own framework. This page is the overview map. It first sets the boundary between a medical device and a wellness product, then lists the compliance building blocks in the order they chain together, with a pointer to the detailed guide for each. The aim is not to explain everything here, but to give you the lay of the land and the sequence.

The first question: medical device or wellness?

Section titled “The first question: medical device or wellness?”

Everything hinges on this boundary, and it is not decided at the sensor but at the level of the claimed intended purpose. Two physically identical watches can fall under different regimes depending on what the manufacturer says about them.

Article 2(1) of MDR (EU) 2017/745 defines a medical device by its medical purpose, diagnosis, prevention, monitoring, prediction, prognosis, treatment or alleviation of disease. If you claim one of these purposes, you are within the scope of the MDR. If you stay on general fitness tracking, you are outside it.

ClaimLikely statusConsequence
Counts steps, estimates calories, tracks sleepWellness (lifestyle)Outside MDR, but radio and battery still apply
Displays a heart rate for information onlyGrey zone, must be documentedDepends on the exact wording of the instructions
Detects atrial fibrillation, alerts the userMedical deviceMDR, class IIa likely, notified body
Measures blood glucose to adjust insulin therapyHigher-risk medical deviceMDR, class IIb to III (up to class III for life-sustaining dosing decisions), reinforced file

The decision is taken early and documented. A medical claim slipped into the marketing of a product designed as a wellness object exposes you to illegal placing on the market. Conversely, over-classifying a lifestyle product needlessly burdens the project. The MDR (EU) 2017/745 guide details qualification and classification.

Once the medical-device status is established, the risk class governs the conformity assessment route, the need for a notified body and the depth of clinical evaluation. For a wearable, the software function is almost always the deciding factor through rule 11 of Annex VIII.

Wearable profileTypical MDR classNotified body
Basic physiological information software with a medical purposeIIaYes
Continuous monitoring whose failure could cause serious harmIIbYes, design dossier examination possible
Device with no measuring function and no decision softwareI (rare for a medical wearable)No, except sub-categories Is, Im, Ir

Plain class I is rare for a wearable with an active medical purpose, because the presence of decision software or a measuring function shifts it to IIa or above. The detail of the 22 rules sits in the MDR guide.

The harmonised technical standards to mobilise

Section titled “The harmonised technical standards to mobilise”

CE marking under MDR rests on the general safety and performance requirements of Annex I. In practice you meet them through a bundle of harmonised standards, each addressing one aspect of the design. No single standard covers a wearable on its own, so the work consists of assembling the right set and showing, document by document, that every applicable clause has been addressed. For a wearable, the base layer is as follows.

The standard IEC 60601-1 covers basic safety and essential performance of medical electrical equipment. Its collateral standard IEC 60601-1-2 deals with electromagnetic compatibility, a sensitive point for a device worn in a home environment saturated with emitters. Usability runs through IEC 60601-1-6 coupled with IEC 62366-1. The IEC 60601-1 medical electrical safety guide goes into the tests and particular standards.

The life cycle of embedded and application software follows IEC 62304, which requires a safety classification (A, B or C) and proportionate activities. Cross-cutting risk management follows ISO 14971, the backbone of the file, which links each identified hazard to a control measure and a verification. See the ISO 14971 risk management guide.

Building blockMain standardWhat it requires
Electrical safetyIEC 60601-1Basic safety and essential performance tests
EMCIEC 60601-1-2Immunity and emissions in a home healthcare environment
UsabilityIEC 60601-1-6 and IEC 62366-1User-centred design, use errors
SoftwareIEC 62304Safety class, life cycle, defect management
RiskISO 14971Analysis, control and acceptability of residual risk
Quality systemISO 13485Quality management dedicated to medical devices

The radio layer, distinct from the medical regime

Section titled “The radio layer, distinct from the medical regime”

Almost every wearable communicates wirelessly, most often over Bluetooth Low Energy, sometimes adding a cellular or LPWAN link for direct connectivity. This radio layer has its own regime that stacks with the MDR, it does not replace it. A common mistake is to assume that medical conformity somehow absorbs the radio requirements, when in fact a wearable that is fully MDR compliant can still be barred from sale for missing a radio approval. Each of the three sub-layers below is assessed and documented on its own track.

In Europe the directive RED (EU) 2014/53 applies, with the standard EN 300 328 for the 2.4 GHz band. In the US, authorisation falls under FCC Part 15 with an FCC ID. The RED checklist guide and the FCC ID, grantee and TCB guide cover these two routes.

A device worn against the body triggers a radiofrequency exposure assessment, the SAR. Measurement follows IEC 62209 and the European exposure recommendation. See the SAR procedures IEC 62209 and EN 50360 guide.

Independently of regulatory compliance, use of the Bluetooth mark requires qualification with the Bluetooth SIG. It is an interoperability certification, not a market authorisation. See the Bluetooth SIG qualification guide.

Selling in the United States adds a whole regime. The most common route for a medical wearable is the FDA 510(k) premarket notification, which demonstrates substantial equivalence to a legally marketed predicate device. With no predicate, the De Novo route applies, and higher-risk devices fall under PMA. The FDA 510(k), De Novo and PMA guide details these routes.

Since 2023, the FDA has required a cybersecurity section in the premarket submission under section 524B of the FD&C Act, applicable to connected devices. On the EU side, cybersecurity sits in Annex I section 17.2 of the MDR, elaborated by guidance MDCG 2019-16. The Cyber Resilience Act (EU) 2024/2847 explicitly excludes medical devices covered by the MDR to avoid a double regime, but it does not reduce the MDR requirements at all. The Cyber Resilience Act (CRA) guide sets out this scope.

A wearable's lithium-ion pack falls under two requirements that do not depend on the medical status. Cell and pack safety follows IEC 62133-2, and air and sea transport falls under the UN Manual of Tests and Criteria section UN 38.3. For the European market, Regulation Battery Regulation (EU) 2023/1542 adds obligations on marking, recycled content and, where applicable, end-user removability. See the battery safety and transport guide.

The sequence below lists the building blocks in a realistic order of work. It is not strictly linear, several streams run in parallel, but MDR qualification governs everything.

  1. Qualify the product, medical device or wellness, and freeze the claim.
  2. Classify the device under Annex VIII of the MDR and choose the conformity route.
  3. Launch ISO 14971 risk management and the ISO 13485 quality system.
  4. Frame the IEC 62304 software life cycle and its safety class.
  5. Plan IEC 60601-1 safety testing and IEC 60601-1-2 EMC.
  6. Document IEC 62366-1 usability.
  7. Freeze the radio hardware, then launch RED or FCC and SAR.
  8. Engage Bluetooth SIG qualification if the mark is used.
  9. Handle the battery, IEC 62133-2, UN 38.3, EU battery regulation.
  10. Integrate cybersecurity, MDCG 2019-16 for the EU, section 524B for the FDA.
  11. Compile the technical file, see the dedicated file-contents guide.
  12. Notified-body audit (EU) then 510(k) submission (US).
MarketFinal documentRegime
European UnionEU declaration of conformity, CE markingMDR plus RED plus battery
United StatesFDA clearance letter, FCC ID510(k) or De Novo plus FCC Part 15
InteroperabilityQualified Design IDBluetooth SIG

The EU declaration of conformity guide and the technical documentation file contents guide close the documentary loop.

PitfallConsequenceCountermeasure
Medical claim in the marketing of a product designed as wellnessIllegal placing on the market under MDRAlign design, instructions and marketing from the start
Software class set too low through an optimistic reading of rule 11Rejection during audit, route changeDocument the classification with an MDCG rationale
Believing Bluetooth SIG qualification equals regulatory complianceProduct non-compliant with RED or FCCTreat the three radio layers separately
Forgetting SAR for a device worn against the bodyHuman-exposure non-compliancePlan the SAR measurement as soon as the radio is frozen
Assuming the CRA replaces the MDR cyber requirementsIncomplete cyber fileApply MDCG 2019-16, the CRA excludes MDR medical devices
Treating the battery as a mere componentTransport or market refusalRequire IEC 62133-2 and UN 38.3 from the pack supplier

Sources & references

  1. Regulation (EU) 2017/745 on medical devices (MDR) , EUR-Lex eur-lex.europa.eu/eli/reg/2017/745/oj
  2. IEC 60601-1, medical electrical equipment, general requirements for basic safety , IEC webstore.iec.ch/publication/2612
  3. IEC 62304, medical device software, software life cycle processes , IEC webstore.iec.ch/publication/22794
  4. ISO 14971, application of risk management to medical devices , ISO www.iso.org/standard/72704.html
  5. FDA, Premarket Notification 510(k) , FDA www.fda.gov/medical-devices/premarket-submissions-selecting-and-preparing-correct-submission/premarket-notification-510k
  6. RED Directive (EU) 2014/53 on radio equipment , EUR-Lex eur-lex.europa.eu/eli/dir/2014/53/oj
  7. MDCG 2019-16, guidance on cybersecurity for medical devices , European Commission health.ec.europa.eu/medical-devices-sector/new-regulations/guidance-mdcg-endorsed-documents-and-other-guidance_en