- All guides
- IEC 61000-4-3: radiated RF field immunity
IEC 61000-4-3: radiated RF field immunity
Radiated RF field immunity is the mirror of radiated emissions: the same kind of absorber-lined room, but the EUT is now the target of a field driven by a power amplifier, and must keep working. IEC 61000-4-3 defines the levels (1, 3, 10 and 30 V/m plus an open level), the 80 % AM test signal, the uniform field area and the sweep. It does not choose the level or the pass criterion: EN 55035, the generic standards and EN 301 489-1 do. The fourth edition dates from 8 September 2020, while the European product and generic standards still refer to the 2006 edition with its two amendments.
Normative frame
Section titled “Normative frame”IEC IEC 61000-4-3 is a basic EMC publication (IEC Guide 107) prepared by IEC SC 77B. The first edition dates from 1995. Edition 4.0 was published on 8 September 2020 and cancels and replaces edition 3.0 (2006), Amendment 1:2007 and Amendment 2:2010. Its stability date on the IEC record is 2027.
The scope is radiated RF fields from sources not in close proximity to the EUT. Since 2020 the scope sends close-proximity sources to IEC 61000-4-39. The standard also states that it is an independent method: other methods cannot be used as substitutes for claiming compliance with it. Product standards can still offer alternatives, which is what EN 55035 does with TEM waveguides (IEC 61000-4-20) and reverberation chambers (IEC 61000-4-21).
Editions and what changed
Section titled “Editions and what changed”| Document | Date | What it means for a test plan |
|---|---|---|
| IEC 61000-4-3 ed. 1.0 | 1995 | First edition |
| IEC 61000-4-3 ed. 3.0 | 2006 | Levels 1 to 4 and X; general range 80 to 1000 MHz; radio telephone ranges 800 to 960 MHz and 1.4 to 6.0 GHz |
| Amendment 1 and Amendment 2 | 2007, 2010 | Folded into edition 4.0 |
| IS 14700 (Part 4/Sec 3) | September 2008 | Indian adoption, identical to IEC 61000-4-3:2006, freely readable |
| IEC 61000-4-21 ed. 2.0 | 2011 | Reverberation chamber method |
| IEC 61000-4-39 ed. 1.0 | 2017 | Radiated fields in close proximity |
| IEC 61000-4-3 ed. 4.0 | 8 September 2020 | Multiple test signals described; more on EUT and cable layout; upper frequency limitation removed; field characterisation and amplifier linearity check specified |
| IEC 61000-4-20 ed. 3.0 | 18 February 2022 | TEM waveguides, including large EUTs and cables |
| EN 55024:2010 | Left the EMC list on 16 September 2022 | EN 55035 is the multimedia immunity route |
| IEC 61000-4-3 stability date | 2027 | Next decision: reconfirm, revise, amend or withdraw |
Which edition the calling standards use
Section titled “Which edition the calling standards use”The calling standards use dated references. EN 55035:2017 (Table 1 note and 4.2.2.2) names IEC 61000-4-3:2006/AMD1:2007/AMD2:2010. ETSI EN 301 489-1 V2.2.3 names EN 61000-4-3 (2006), A1 (2008) and A2 (2010), clauses 6, 7 and 8. The generic standards IEC 61000-6-1:2005 and IEC 61000-6-2:2005 cite IEC 61000-4-3 without a date. For a test claimed against these standards, the 2006 text with its amendments is the reference, which is why the clause details below are quoted from that edition. In edition 4.0 the clauses moved: test levels and frequency ranges are clause 5, the uniform field area is 6.3, the test setup is clause 7, step sizes are 8.4 and evaluation is clause 9.
What the calling standards ask
Section titled “What the calling standards ask”| Calling standard (EU list status) | Frequency range | Level (unmodulated) | Modulation | Criterion |
|---|---|---|---|---|
| EN 55035:2017+A11:2020, multimedia (cited) | 80 to 1000 MHz swept | 3 V/m | 80 % AM, preferably 1 kHz | A |
| EN 55035, spot frequencies (±1 %) | 1800, 2600, 3500, 5000 MHz | 3 V/m | 80 % AM, preferably 1 kHz | A |
| EN 61000-6-1:2007, residential, commercial, light industry (cited) | 80 to 1000 MHz; 1.4 to 2.0 GHz; 2.0 to 2.7 GHz | 3 V/m; 3 V/m; 1 V/m | 80 % AM, 1 kHz | A |
| EN 61000-6-2:2005, industrial (cited) | 80 to 1000 MHz; 1.4 to 2.0 GHz; 2.0 to 2.7 GHz | 10 V/m (3 V/m in 87 to 108, 174 to 230 and 470 to 790 MHz); 3 V/m; 1 V/m | 80 % AM, 1 kHz | A |
| ETSI EN 301 489-1 V2.2.3, radio equipment, clause 9.2 | 80 to 6000 MHz, outside the exclusion band | 3 V/m | 80 % AM, 1 kHz (400 Hz if the wanted signal uses 1 kHz) | A (continuous phenomena) |
| IEC 60601-1-2 ed. 4.1, medical, Table 4 | 80 MHz to 2.7 GHz | 3 V/m professional healthcare facility; 10 V/m home healthcare | 80 % AM, 1 kHz | Basic safety and essential performance |
| ISO 11452-2:2019, road vehicle components | 80 MHz to 18 GHz | Specified by the user; suggested levels in Annex B | Per ISO 11452-1 | Per test plan |
EN 55035 supersedes EN 55020, EN 55103-2 and EN 55024; its date of withdrawal for conflicting standards was 28 July 2022. The 2019 editions of the generic standards are not in the EMC list (see CE tests). For radio equipment, the level and range come from EN 301 489-1 (see RED tests).
EN 55035 adds three options that IEC 61000-4-3 does not give by itself: the swept test may use a step of up to 4 % at twice the specified level, testing may be restricted to the most sensitive face if it is known over the whole range, and the independent windows method (0.5 m by 0.5 m) may be used for the spot frequencies above 1 GHz.
The test signal
Section titled “The test signal”The test level is the RMS value of the unmodulated carrier. For testing, the carrier is 80 % amplitude modulated by a 1 kHz sine wave. With 80 % AM the peak-to-peak envelope reaches 1.8 times the carrier: for a 1 V carrier, the maximum RMS value defined in the standard is 1.8 V, which is 20 log(1.8) = 5.1 dB above the carrier.
Annex A gives the reasoning against pulsed modulations. Tests compared sine wave 80 % AM at 1 kHz, a "GSM-like" 1:8 duty cycle at 200 Hz and "DECT-like" pulsing at 100 Hz on hearing aids, analogue telephones and radio sets. The conclusion was that the immunity level correlates with the maximum RMS value of the field, and sine wave AM was kept because it is simple, readily generated, and already used in IEC 61000-4-6. Its listed disadvantages are that it does not simulate TDMA and may miss some failure mechanisms.
Severity levels and environment classes
Section titled “Severity levels and environment classes”| Level | Carrier field | Maximum RMS with 80 % AM | Environment class in Annex E | Protection distance, 2 W GSM | 8 W GSM | 0.25 W DECT |
|---|---|---|---|---|---|---|
| 1 | 1 V/m | 1.8 V/m | Low level: broadcast stations beyond 1 km, low-power transmitters | 5.5 m | 11 m | 1.9 m |
| 2 | 3 V/m | 5.4 V/m | Moderate: portable transceivers under about 1 W with restrictions; typical commercial | 1.8 m | 3.7 m | 0.6 m |
| 3 | 10 V/m | 18 V/m | Severe: portable transceivers of 2 W or more, not closer than 1 m; typical industrial | 0.6 m | 1.1 m | 0.2 m |
| 4 | 30 V/m | 54 V/m | Portable transceivers within 1 m | 0.2 m | 0.4 m | 0.1 m |
| X | Open | Negotiated in the product standard or specification |
The protection distances come from Table E.1 of IEC 61000-4-3:2006 (Equation E.1, E = k√P / d with k = 7, using the maximum RMS field and the effective radiated power). The standard warns that the far-field equation is not accurate at the shortest distances and that reflections cause fluctuations of the order of ±6 dB. The table is a planning aid: it tells you how close a phone can come to equipment tested at a given level.
Test facility and field calibration
Section titled “Test facility and field calibration”Facility and equipment
Section titled “Facility and equipment”Because of the field strengths involved, the test is made in a shielded enclosure. The facility is typically an absorber-lined room: a fully anechoic chamber, or a semi-anechoic chamber modified with additional absorbers on the ground plane. Recommended equipment is an RF generator able to apply 80 % AM at 1 kHz, a power amplifier whose harmonics in the UFA stay at least 6 dB below the fundamental, linearly polarised antennas (biconical, log-periodic, horn or other), an isotropic field probe with a fibre-optic link, and means of recording the forward power. The 2006 measuring setup chains a controller, signal generator, power amplifier, directional coupler and power meter, transmitting antenna, field sensor and field meter.
Controller -> Signal generator -> Power amplifier -> Directional coupler -> Transmitting antenna | | Power meter field in the UFA | Field meter <- fibre <- Isotropic field sensor (calibration, no EUT)Annex C notes that ferrite tiles usually behave as reflectors above 1 GHz, so a small ferrite-lined chamber may fail the uniformity requirement there. The suggested fixes are a horn or double-ridge antenna, a shorter distance (down to 1 m) with the independent windows method, and carbon-loaded absorber on the wall behind the EUT. See EMC chamber types.
Uniform field area
Section titled “Uniform field area”| UFA parameter (IEC 61000-4-3:2006, 6.2) | Requirement |
|---|---|
| Plane | Vertical, calibrated with no EUT in place, both polarisations |
| Size | At least 1.5 m by 1.5 m unless the EUT and its wires fit a smaller area; never below 0.5 m by 0.5 m |
| Lower edge | 0.8 m above the floor; field also recorded at 0.4 m for information |
| Grid | 0.5 m spacing (16 points for 1.5 m by 1.5 m) |
| Tolerance | -0 dB to +6 dB of the nominal value |
| Compliant points | At least 75 % (12 of 16); all 4 points for a 0.5 m by 0.5 m UFA |
| Distance | Field sensor at least 1 m from the antenna; 3 m preferred; distance stated in the report |
| Frequency step | At most 1 % of the present frequency |
| Calibration field | Ec at least 1.8 times the test field Et, with an unmodulated carrier |
| Saturation check | Reduce the generator by 5.1 dB: the forward power must drop by 3.1 to 5.1 dB, otherwise the amplifier is saturated |
| Validity | Full calibration annually and after chamber changes; validity checked before each batch of tests |
Two calibration methods are given, constant field strength and constant power, and they are considered equivalent. The calibration records the forward power for the calibration field. During the test, the forward power is calculated from it: Pt = Pc - 20 log(Ec / Et). In edition 4.0 the UFA is clause 6.3, a minimum-size UFA has a fifth grid point in the centre (Figure 5), and the amplifier linearity check is specified in Annex D.
When the EUT face does not fit the UFA, partial illumination (with a UFA of at least 1.5 m by 1.5 m) or, above 1 GHz, the independent windows method is used.
Test setup and procedure
Section titled “Test setup and procedure”EUT and cable layout
Section titled “EUT and cable layout”- Configuration as close as possible to the actual installation, with covers in place; panel or rack equipment tested in its panel or rack.
- No metallic ground plane is required. Table-top equipment on a non-conductive table 0.8 m high; floor-standing equipment on a non-conductive support 0.05 m to 0.15 m high.
- Low-permittivity supports such as rigid polystyrene, because wood or glass-reinforced plastic can reflect above 1 GHz.
- Manufacturer-specified cables; unshielded parallel conductors when nothing is specified. If possible, at least 1 m of cable exposed to the field. Excess length of cables between units of the EUT bundled low-inductively at the approximate centre into a bundle 30 cm to 40 cm long.
- Body-worn equipment may be tested like table-top equipment, but product committees are encouraged to specify a human body simulator.
Edition 4.0 adds informative examples for bottom-fed and overhead cables, multiple cables and auxiliary equipment, and large EUTs with several UFA windows, and defines a common-mode absorption device that may be applied on cables leaving the test area.
Procedure
Section titled “Procedure”- Write the test plan: EUT size and operating conditions, table-top or floor-standing, facility and antennas, frequency range, dwell time and step, UFA size, partial illumination, test level, cables and ports, acceptable performance criteria, and how the EUT is exercised.
- Verify the calibrated field before testing.
- Place the EUT with one face in the calibration plane.
- Sweep with the modulated signal. In incremental sweeps the step does not exceed 1 % of the preceding frequency.
- Dwell at each frequency for at least the time needed to exercise the EUT and let it respond, and never less than 0.5 s. Sensitive frequencies such as clock frequencies are analysed separately as the product standard requires.
- Normally test each side of the EUT, each with the antenna vertical and then horizontal. Fewer faces may be justified; large EUTs may need more than four azimuths.
- Record effects and their duration, and classify them.
Performance criteria
Section titled “Performance criteria”IEC 61000-4-3 recommends a classification, not a verdict:
| Class (clause 9) | Observed effect |
|---|---|
| a | Normal performance within the specified limits |
| b | Temporary loss of function or degradation that ceases with the disturbance, recovered without operator intervention |
| c | Temporary loss of function or degradation that needs operator intervention |
| d | Loss of function or degradation not recoverable, due to damage to hardware or software, or loss of data |
The generic and product standards turn this into criteria A, B and C. For continuous radiated RF, every calling standard in the table above applies criterion A: EN 55035 defines it as continuing to operate as intended without operator intervention, with no degradation, loss of function or change of operating state below the performance level specified by the manufacturer. IEC 60601-1-2 instead assesses basic safety and essential performance (see IEC 60601-1).
Medical proximity fields
Section titled “Medical proximity fields”IEC 60601-1-2 ed. 4.1 adds proximity tests with RF wireless communication equipment on top of the 80 MHz to 2.7 GHz sweep. As reproduced in manufacturer declarations, Table 9 covers seven bands: 380 to 390 MHz at 27 V/m, 430 to 470 MHz at 28 V/m, 704 to 787 MHz at 9 V/m, 800 to 960 MHz at 28 V/m, 1700 to 1990 MHz at 28 V/m, 2400 to 2570 MHz at 28 V/m and 5100 to 5800 MHz at 9 V/m, with pulse modulation at 18 Hz or 217 Hz (FM is also listed for 430 to 470 MHz). The proximity magnetic field tests use IEC 61000-4-39. For proximity sources in general, IEC 61000-4-3:2020 points to IEC 61000-4-39.
Claims the texts contradict
Section titled “Claims the texts contradict”| Common claim (including on earlier versions of this page) | What the text says |
|---|---|
| Edition 4.0 (2020) added a "1.4 to 6 GHz extension" | The 2006 edition already tested 800 to 960 MHz and 1.4 to 6.0 GHz for digital radio telephones (clause 5.2). The 2020 edition removed the upper frequency limitation |
| The levels are 3, 10 and 20 V/m | Table 1 gives 1, 3, 10 and 30 V/m (levels 1 to 4) and an open level X; 20 V/m is not a listed level |
| "Level X is 30 V/m and above" | 30 V/m is level 4; X is an open level of any value |
| EN 55024 is a current route for IT equipment | EN 55024:2010 left the EMC list on 16 September 2022; EN 55035 supersedes it |
| EN 55035 asks 3 V/m only from 80 to 1000 MHz | It also asks 3 V/m at 1800, 2600, 3500 and 5000 MHz (±1 %) |
| EN 61000-6-2 asks 10 V/m everywhere from 80 to 1000 MHz | 3 V/m in the broadcast bands 87 to 108, 174 to 230 and 470 to 790 MHz |
| IEC 61000-4-3 does not deal with performance at all | Clause 9 recommends a four-step classification (a to d); the criteria A, B, C come from the product or generic standard |
| Medical: 10 V/m for "urgent care" | IEC 60601-1-2 Table 4 uses 3 V/m for the professional healthcare facility and 10 V/m for the home healthcare environment |
| I/O cables are laid in a 30 to 40 cm serpentine | At least 1 m of cable is exposed if possible; only excess length between EUT units is bundled, low-inductively, into a 30 to 40 cm bundle |
| A 3 V/m calibration cannot be scaled to another level because the field distribution changes | The test power is derived from the calibration as Pt = Pc - 20 log(Ec / Et). What limits it is amplifier saturation: Et must not exceed Ec / 1.8 |
| Dwell time 0.5 to 3 s | At least 0.5 s and at least the time to exercise the EUT and let it respond; EN 55035 says it should not exceed 5 s |
| ISO 11452-2 goes "up to 200 V/m", "level IV 100 V/m, up to 2.7 GHz" | ISO 11452-2:2019 covers 80 MHz to 18 GHz; the user specifies the severity, with suggested levels in an informative annex |
Mitigating a radiated immunity failure
Section titled “Mitigating a radiated immunity failure”When the EUT fails at a given frequency, the approach is the inverse of emissions: reduce coupling of the field into the sensitive internal point. Three paths are common.
- Cable coupling. The cable acts as a receiving antenna and brings a common-mode current to the circuit. Mitigation: common-mode ferrite effective at the failing frequency, filtering at the connector referenced to the chassis, shield terminated 360 degrees.
- Aperture coupling. A slot or ventilation opening radiates efficiently when its length approaches half a wavelength. Mitigation: shorter openings, conductive gaskets on removable covers, mesh on vents.
- Direct PCB coupling. At higher frequencies, traces and high-impedance nodes can pick up the field directly. Mitigation: local shielding, decoupling, continuous ground plane, guard traces on sensitive analogue signals.
Identify the path with discriminating experiments: add a ferrite on one cable, close one aperture temporarily with copper tape, shield one PCB area. See EMC coupling mechanisms.
What to remember
Section titled “What to remember”- IEC 61000-4-3 ed. 4.0 was published on 8 September 2020 and replaces the 2006 edition and its 2007 and 2010 amendments. EN 55035:2017, EN 301 489-1 V2.2.3 and the generic standards still use dated references to the 2006 text.
- Levels are 1, 3, 10 and 30 V/m plus X. The level is the unmodulated carrier; 80 % AM at 1 kHz brings the maximum RMS to 1.8 times that value.
- The UFA is a vertical plane with a 0.5 m grid, normally 1.5 m by 1.5 m, with -0/+6 dB at 75 % of the points, calibrated without the EUT at 1.8 times the test field.
- The sweep uses steps of at most 1 % and a dwell of at least 0.5 s, on each face in both polarisations.
- EU levels: 3 V/m and spot frequencies to 5 GHz for multimedia (EN 55035), 3 or 10 V/m for the generic standards with 1.4 to 2.7 GHz bands, 3 V/m to 6 GHz for radio equipment (EN 301 489-1). Criterion A in all of them.
For EU application see CE tests and RED tests. Acronyms are in the glossary.
See also
Section titled “See also”- Radio: RX blocking, selectivity and intermodulation tests
- IEC 61000-4-6: conducted RF immunity
- IEC 61000-4-8: power-frequency magnetic field immunity
- ESD per IEC 61000-4-2: method and test levels
- Surge and Burst (IEC 61000-4-5 / 4-4): EMC transients
- Voltage dips and interruptions (IEC 61000-4-11)
- Radiated emissions
- EMC chamber types
- ECE R10 automotive EMC
Sources & references
- IEC 61000-4-3:2020 (Edition 4.0), published 8 September 2020, stability date 2027, significant technical changes, IEC webstore.iec.ch/en/publication/59849
- IEC 61000-4-3:2020 preview: foreword, scope, terms, table of contents, IEC (preview hosted by iTeh Standards) cdn.standards.iteh.ai/samples/100266/3b514021a3db4adbbf46899a64b655e3/IEC-61000-4-3-2020.pdf
- IEC 61000-4-3:2006 (Edition 3.0) record, IEC webstore.iec.ch/publication/4209
- IS 14700 (Part 4/Sec 3):2008, identical to IEC 61000-4-3:2006, full text (Table 1, clauses 6.2, 7, 8, 9, Annexes A, C, E), Bureau of Indian Standards (via Public.Resource.Org) law.resource.org/pub/in/bis/S04/is.14700.4.3.2008.pdf
- EN 55035:2017+A11:2020, Electromagnetic compatibility of multimedia equipment, immunity requirements, CENELEC www.cenelec.eu/dyn/www/f?p=104:110:::::FSP_PROJECT,FSP_LANG_ID:60147,25
- CISPR 35:2016, electromagnetic compatibility of multimedia equipment, immunity requirements, IEC webstore.iec.ch/en/publication/25667
- Commission Implementing Decision (EU) 2019/1326, harmonised standards for EMC, consolidated text of 10 June 2022, Publications Office of the European Union publications.europa.eu/resource/celex/02019D1326-20220610
- Commission communication under Directive 2014/30/EU, OJ C 246, 13 July 2018 (EN 61000-6-1:2007, EN 61000-6-2:2005), Publications Office of the European Union publications.europa.eu/resource/celex/52018XC0713%2802%29
- IS 14700 (Part 6/Sec 1):2008, identical to IEC 61000-6-1:2005, Table 1 enclosure port, Bureau of Indian Standards (via Public.Resource.Org) law.resource.org/pub/in/bis/S04/is.14700.6.1.2008.pdf
- IS 14700 (Part 6/Sec 2):2008, identical to IEC 61000-6-2:2005, Table 1 enclosure port, Bureau of Indian Standards (via Public.Resource.Org) law.resource.org/pub/in/bis/S04/is.14700.6.2.2008.pdf
- ETSI EN 301 489-1 V2.2.3 (2019-11), clause 9.2, radio frequency electromagnetic field (80 MHz to 6 000 MHz), ETSI www.etsi.org/deliver/etsi_en/301400_301499/30148901/02.02.03_60/en_30148901v020203p.pdf
- IEC 61000-4-20:2022 (Edition 3.0), emission and immunity testing in TEM waveguides, IEC webstore.iec.ch/en/publication/34378
- IEC 61000-4-21:2011 (Edition 2.0), reverberation chamber test methods, IEC webstore.iec.ch/en/publication/4191
- IEC 61000-4-39:2017, radiated fields in close proximity, immunity test, IEC webstore.iec.ch/en/publication/30202
- IEC 60601-1-2:2014+AMD1:2020 CSV, Medical electrical equipment, collateral standard, electromagnetic disturbances, IEC webstore.iec.ch/publication/67554
- Manufacturer EMC declaration reproducing the IEC 60601-1-2 ed. 4.1 radiated and proximity test levels (secondary source), PARI Pharma www.pari.com/fileadmin/Electromagnetic-compatibility-4.pdf
- ISO 11452-2:2019, Road vehicles, component test methods for electrical disturbances from narrowband radiated electromagnetic energy, absorber-lined shielded enclosure, ISO www.iso.org/standard/72840.html
Frequently asked questions
- How does IEC 61000-4-3 differ from a radiated emissions measurement?
- Both use an absorber-lined shielded room and antennas, but the roles are reversed. In radiated emissions (CISPR 32, CISPR 11) the EUT is the source and the antenna receives. In IEC 61000-4-3 a signal generator and power amplifier drive the antenna, the EUT sits in a field set from a prior calibration, and the test looks for loss of function. See the Radiated emissions guide for the emissions counterpart.
- Why 80 % AM at 1 kHz?
- The test field strength is the unmodulated carrier, then modulated 80 % by a 1 kHz sine wave. Annex A of IEC 61000-4-3 explains the choice: experiments showed good correlation between the effects of different non-constant envelope modulations as long as the maximum RMS level stays the same, the equipment is readily available, and the same modulation is used in IEC 61000-4-6. With 80 % AM the maximum RMS value is 1.8 times the carrier, that is 5.1 dB above it. EN 301 489-1 switches to 400 Hz when the wanted signal is itself modulated at 1 kHz.
- What field level applies to what product?
- IEC 61000-4-3 lists levels 1, 2, 3 and 4 at 1, 3, 10 and 30 V/m plus an open level X, and leaves the choice to product committees. In the EU lists, EN 55035 (multimedia) asks for 3 V/m from 80 to 1000 MHz plus spot frequencies 1800, 2600, 3500 and 5000 MHz. EN 61000-6-1:2007 asks for 3 V/m, EN 61000-6-2:2005 for 10 V/m (3 V/m in broadcast bands) from 80 to 1000 MHz, both with 3 V/m from 1.4 to 2.0 GHz and 1 V/m from 2.0 to 2.7 GHz. EN 301 489-1 asks for 3 V/m from 80 MHz to 6 GHz for radio equipment.
- What is the uniform field area?
- The UFA is a vertical plane in which the field is calibrated without the EUT. It is subdivided into a 0.5 m grid, normally 1.5 m by 1.5 m (16 points) with its lower edge 0.8 m above the floor. At each frequency the field must lie within -0 dB to +6 dB of the nominal value at no fewer than 75 % of the points, that is 12 of 16. The UFA may be smaller if the EUT and its wiring fit, but not smaller than 0.5 m by 0.5 m, where all points must comply.
- Who defines criteria A, B and C?
- Not IEC 61000-4-3. Its clause on evaluation only recommends a four-step classification (normal performance, temporary loss with self-recovery, loss needing operator intervention, unrecoverable loss). The product or generic standard sets the criterion. For continuous radiated RF, EN 55035, EN 61000-6-1, EN 61000-6-2 and EN 301 489-1 all apply criterion A.
- Does IEC 61000-4-3 stop at 6 GHz?
- Edition 3.0 (2006) normally tested 80 to 1000 MHz for general purposes and 800 to 960 MHz plus 1.4 to 6.0 GHz for digital radio telephones. Edition 4.0 (2020) removed the upper frequency limitation to take account of new services. The 1.4 to 6 GHz range is therefore not a new extension of the 2020 edition, and the upper limit actually tested comes from the product standard.
- How long does an 80 MHz to 1 GHz sweep take?
- With steps of 1 % of the preceding frequency, 80 to 1000 MHz is about 254 frequencies. The dwell time is at least 0.5 s and EN 55035 says it should not exceed 5 s, so one face in one polarisation takes from about 2 to 21 minutes, before level changes and antenna switching. Four faces in two polarisations multiply that by eight. EN 55035 also allows a 4 % step at twice the test level, about 65 frequencies over the same range.
- What if the EUT fails at a single frequency?
- Record the frequency, polarisation, face and observed effect, then check that it repeats with the same cable layout and operating mode. Then find the coupling path with discriminating experiments: a ferrite on one cable, copper tape over one aperture, a local shield over one PCB area. The report must state any special conditions of use, such as cable type or shielding, that were needed to achieve compliance.